Cygnus 3 - Datalogger
Multiple Echo Ultrasonic Thickness Gauge
The Cygnus 3 Data Logger is versatile and easy to program with simple menu operation. Three separate formats are used to collect measurements and provide solutions to recording various geometries such as tanks, cylinders, pipes and structures.
|
- Simple Linear Data Logging mode with a capacity of up to 5000 points in each record
- Grid Based Data Logging mode uses simple xy grids for up to 5000 points
- Template based data logging mode features definable measurement points, minimum & reference thickness values and up to 40 configurable user fields to collect additional information
- Facility to add Additional Radial Measurements around any chosen point
- Alphanumeric data entry via phone style keys
- Auto-log function automatically logs stable thickness measurements for one-handed operation
- Data stored on internal SD memory card for security
- Velocity Lookup Table can be customized by the user
- USB connection to computer without driver installation
- Supplied with - Data Logger Manager Software for generating reports and transferring data between the gauge and computer - Microsoft Excel add-in to easily import data into blank worksheets or Excel templates
- Housed in a rugged, IP65 rated aluminum enclosure
- Gauge senses probe type and automatically adjusts setting for optimum performance
- LCD graphic display with automatic white backlight
- Simple gauge and menu operation
- Secure, twist-to-lock probe connection (BNC)
- Deep-Coat mode for coatings up to 20mm thick
- Protective silicone sleeve offers maximum protection
- Rugged probe cable designed for superior flexibility
|
|
| Materials |
Sound velocities between 0.059 to 0.31 in/microsec (2,000 m/s to 7,000 m/s)
|
| Range in Steel |
0.110" - to 9.995" (3mm250mm) with 2.25MHz probe 0.065" - to 6.000" (2mm150mm) with 3.5MHz probe 0.045" - to 4.000" (1mm50mm) with 5.0MHz probe
|
| Accuracy |
+/- 0.005" (0.1mm) or +/- 0.002" (0.05mm) |
| Resolution |
0.005" (0.1mm) or 0.002" (0.05mm) |
| Probes |
Single crystal soft-faced compression 1/2" standard (13mm) - 2.25 MHz 1/4" (6mm) - 5.0 MHz 1/2" (13mm) - 5.0 MHz or 3.5 MHz 3/4" (19mm) - 2.25 MHz
|
| Power |
3 x AA Alkaline batteries or rechargeable NiMH / NiCD |
| Battery Life |
18 hours continuous operation with alkaline batteries |
| Display |
Large liquid crystal display with white backlightautomatically turns off in bright light conditions |
| Size |
3.34 in x 6.69 in x 0.98 in (85mm x 170mm x 25mm) |
| Operating Temp |
14 F - 122 F (-10C50C) |
| Data Output Storage |
Internal storage on Flash memory card. Data output via USB connection |
| Data Logger (Maximum) |
Measurement points per record: 5000 Records: 100 Grid record size: 100 rows x 50 columns User data fields: 40
|
| Environmental protection |
IP65 rated | |
Instrument, 2.25MHz 1/2" probe, operations manual, carry case, protective silicone sleeve, adjustable neck strap, accessory pouch (containing couplant gel, spare membranes, membrane couplant, membrane locking ring key, 1/2" test block, 3 x AAA batteries), software, USB cable, probe cable and belt/harness clip.

THREE YEAR WARRANTY
|
|
- Metal Thickness measurements on cranes, marine structures and conveying systems
- Corrosion checks on ships’ shell plates, bulkheads and structures
- Metal thickness safety checks on steam and pressurized water systems,
transportable gas containers and compressed air systems.
- Systematic wall thickness and corrosion monitoring of storage tanks, process vessels and road tankers.
- Quality assurance metal thickness checks.
- Maintenance and safety checks on bridges and street lighting columns.
- Pipeline wall thickness monitoring.
|
| |
|
|